No Cover Image

Journal article 432 views 64 downloads

A Monte Carlo model of an agility head for a 10-MV photon beam

Turki Almatani, Richard Hugtenburg Orcid Logo, Artjoms Smakovs

Journal of Taibah University for Science, Volume: 16, Issue: 1, Pages: 300 - 307

Swansea University Author: Richard Hugtenburg Orcid Logo

  • 62051.pdf

    PDF | Version of Record

    © 2022 The Author(s). This is an Open Access article distributed under the terms of the Creative Commons Attribution License

    Download (1.55MB)

Abstract

Objective: Modelling linear accelerator (linac) photon beams using the Monte Carlo (MC) technique is presently the most precise technique with which to calculate radiotherapy beam data. Complicated treatment plans can be achieved by using the Agility multileaf collimator. The focus of this study is...

Full description

Published in: Journal of Taibah University for Science
ISSN: 1658-3655
Published: Informa UK Limited 2022
Online Access: Check full text

URI: https://cronfa.swan.ac.uk/Record/cronfa62051
Tags: Add Tag
No Tags, Be the first to tag this record!
Abstract: Objective: Modelling linear accelerator (linac) photon beams using the Monte Carlo (MC) technique is presently the most precise technique with which to calculate radiotherapy beam data. Complicated treatment plans can be achieved by using the Agility multileaf collimator. The focus of this study is to model a linac, integrated with an Agility head, using EGSnrc MC code and to investigate its accuracy compared with actual measurements. Methodology: A 10-MV photon beam from an Agility head installed in an Elekta Synergy linac was simulated using BEAMnrc MC code. Various incident electron energies were used to generate percentage depth dose curves for the MC model and were compared with measurements using gamma analysis (γ). Consequently, the focal spot with different full width half-maximums (FWHMs) was used to tune dose profiles. In addition, any differences in dose profiles were further minimized using different angular divergences. Results: The electron energy of 9.6 MeV was found to represent the best match with the measurements, where the γ pass rate was 100%. For dose profile comparisons, a circular focal spot with an FWHM of 0.45 cm and an angular divergence of 0.04∘ produced the best match with the measurements. Conclusion: The presented MC linac model integrated with an Agility head for a 10-MV photon beam can be reliably used for dose calculations.
Keywords: Agility head; MC; BEAMnrc;DOSXYZnrc
College: Faculty of Medicine, Health and Life Sciences
Issue: 1
Start Page: 300
End Page: 307