Journal article 1838 views 509 downloads
Study of Electrochemical Impedance of a Continuous Glucose Monitoring Sensor and its Correlation with Sensor Performance
IEEE Sensors Letters, Pages: 1 - 1
Swansea University Author:
Sanjiv Sharma
-
PDF | Accepted Manuscript
Download (577.72KB)
DOI (Published version): 10.1109/LSENS.2017.2778248
Abstract
In this work, we study the change in the sensitivity and the electrochemical impedance of continuous glucose monitoring sensors over time. 28-day sensitivity and EIS measurement results on four similar sensors are presented. The sensitivity of the sensor is observed to be related to its double-layer...
| Published in: | IEEE Sensors Letters |
|---|---|
| ISSN: | 2475-1472 |
| Published: |
2017
|
| Online Access: |
Check full text
|
| URI: | https://cronfa.swan.ac.uk/Record/cronfa37061 |
| Abstract: |
In this work, we study the change in the sensitivity and the electrochemical impedance of continuous glucose monitoring sensors over time. 28-day sensitivity and EIS measurement results on four similar sensors are presented. The sensitivity of the sensor is observed to be related to its double-layer capacitance and charge-transfer resistance, based on results acquired from a sensor that showed substantial sensitivity drop. Two data clusters are extracted that relate the sensor sensitivity to its impedance before and after the sensitivity drops by more than 50%. |
|---|---|
| Keywords: |
Impedance spectroscopy, glucose sensor, sensitivity, correlation, double-layer capacitance |
| College: |
Faculty of Science and Engineering |
| Start Page: |
1 |
| End Page: |
1 |